WOW !! MUCH LOVE ! SO WORLD PEACE !
Fond bitcoin pour l'amélioration du site: 1memzGeKS7CB3ECNkzSn2qHwxU6NZoJ8o
  Dogecoin (tips/pourboires): DCLoo9Dd4qECqpMLurdgGnaoqbftj16Nvp


Home | Publier un mémoire | Une page au hasard

 > 

Infinite dilution activity coefficient measurements of organic solutes in fluorinated ionic liquids by gas-liquid chromatography and the inert gas stripping method

( Télécharger le fichier original )
par Kaniki TUMBA
University of Kwazalu-Natal - Master 2009
  

précédent sommaire suivant

Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy

4.2.2.4. Cold trap

The stream leaving the dilutor cell contains not only nitrogen, but also solute and probably solvent vapours. A cold trap serves as a separation unit to ensure that the flow rate determined by the soap bubble flow-meter is actually the one of nitrogen which is required in the infinite dilution activity coefficient computation equation. Failure to condense solute and solvent vapours will lead to inaccurate data. The cold trap as shown in figure 4-2 consists of two chambers. The upper chamber (A) contains an ice-acetone solution in which a 1/4 inch ID copper coil is completely immersed. The outlet gas stream from the dilutor cell flows through the coils and ends its path in the lower chamber (B) where the condensate is trapped. Nitrogen, the only gaseous component after this process, rises up to exit through the pipe (C) and makes its way to the soap bubble flow-meter (J). Valve (E) can be opened to drain out the liquid captured in chamber (B).

Photograph 4-2: Set-up of the inert gas stripping apparatus. A-flow regulator; B-Coil tube
(Heat exchanger); C-Presaturation cell; D- Dilutor cell; E-Magnetic stirrer; F-Sampling valve;
G-Cold trap; H-GC apparatus; I-GC lines; J-Soap bubble flow-meter; K-Pressure transducer; L-
Pressure and temperatures displays; M-Immersion temperature controller; N-Polystyrene chips;
O-Transparent acrylic bath.

Photograph 4.3: The dilutor cell. A-Glass still; B-Capillaries; C-Temperature probe (Pt 100);
D-Gas inlet; E-Gas outlet; F-Pressure transducer tube; G-Bolt (seal); H-Teflon plug (seal).

Figure 4-2: Cross Section of the cold trap to illustrate its inner workings (George 2008)
A - Upper chamber, B - lower chamber, C - gas inlet, D - non-condensable gas outlet, E -
release valve, F - lid.

4.2.2.5. Gas Chromatography apparatus

In this work, a SHIMADZU GC 14A gas chromatograph equipped with a FID detector was used. The FID has the advantage over the TCD to be very sensitive to small amounts of solutes injected. GC settings and column conditions during experiments are displayed in table 4-1. To avoid condensation of the solute as the stream flows from the dilutor cell outlet to the GC injection port via the six-port sampling valve, these lines had to be heated and thermostated at a temperature approximately 40 oC higher than the system temperature (Krummen et al. 2000).

Table 4-1: GC specification and set-up.

Column CRS; 2m x 1/8`; Packed column

GC Program Clarity work station

Detector type FID

Carrier gas Helium

Injection Port Temperature 493.15 K

Column Oven Temperature 393.15 K-453.15 K

Detector Temperature 493.15 K

4.2.3. Experimental procedure

First, the cells were cleaned and thereafter filled with the required components. The solvent was poured into each cell. The mass of the solvent in the dilutor cell had to be known accurately. The solute was injected into the dilutor cell in very small concentration to comply with the infinite dilution requirement, less than 0.001 mole fraction. The injected volume ranged between (10 and 25). Density data of the solvent (given in appendix C) were used to

determine by difference the vapour phase volume. The cells were then sealed and fitted onto the set up. The equipment was checked for leaks, the cold trap upper chamber was filled with acetone-ice mixture, the GC apparatus and all electrical equipments with desired settings were switched on. Thereafter the procedure below was followed:

1. The water bath, the sampling valve, the mixture in the dilutor cell and all other tubing were allowed to equilibrate to their respective set-point temperatures, the magnetic stirrer being switched on;

2. The inert gas was allowed to flow through the cell. A bubble soap flow-meter was used to measure the flow rate. Li et al. (1993) suggested that the infinite dilution activity coefficient obtained should be independent of the flow rate;

3. The gas sampling valve was set to the fill? position, allowing the inert gas to flow for some minutes;

4. The gas sampling valve was set to the inject? position for one to two minutes;

5. The sampling valve was set back to the fill? position;

6. Steps 4 and 5 were repeated periodically. The time interval between two successive injections depended on the system under investigation. The natural logarithm of the solute peak area was expected to decrease linearly with time as illustrated by figure 4-3 which was taken from a publication by Krummen et al. (2000);

7. After a number of runs with sufficient data to compute infinite dilution activity coefficient values, the experiment was ended. Equation (3-90) was used to determine experimental IDAC values.

Figure 4-3: Typical plots of solute GC peak area and ln (solute peak area) versus time.

précédent sommaire suivant






Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy








"Le don sans la technique n'est qu'une maladie"