WOW !! MUCH LOVE ! SO WORLD PEACE !
Fond bitcoin pour l'amélioration du site: 1memzGeKS7CB3ECNkzSn2qHwxU6NZoJ8o
  Dogecoin (tips/pourboires): DCLoo9Dd4qECqpMLurdgGnaoqbftj16Nvp


Home | Publier un mémoire | Une page au hasard

 > 

Infinite dilution activity coefficient measurements of organic solutes in fluorinated ionic liquids by gas-liquid chromatography and the inert gas stripping method

( Télécharger le fichier original )
par Kaniki TUMBA
University of Kwazalu-Natal - Master 2009
  

précédent sommaire suivant

Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy

4.1.2. Experimental set up

Experiments were performed using a Shimadzu GC-2014 gas chromatograph apparatus, equipped with a thermal conductivity detector, an auto-sampler and auto-injector. Retention times and chromatograms related to each run were made available on a PC monitor by means of a GC solution workstation software. The set up, shown in Photograph 4-1, is programmable so that it can carry out injections of samples from 125 different vials and acquire data with minimum human supervision.

Photograph 4-1: Gas-Liquid Chromatography equipment.
1. Vials; 2. Tray; 3. Auto-sampler (arm); 4. Column oven; 5. Soap bubble flow-meter; 6. Auto-
injector; 7. GC screen; 8. Operation panel; 9. PC monitor

4.1.3. Experimental procedure

The experimental procedure used in this work has been well documented by numerous authors (David et al. 2003, Letcher et al. 2003a,b and Deenadayalu et al. 2005). To check for adsorption problems, two different stainless steel columns of length 1 m and 4.1 mm inner diameter were used with two different mass percent packings. To avoid adsorption effects, mass percent packings of the ionic liquid that are large enough were used. Before packing, the columns were washed with hot soapy water, rinsed with cold water and flushed with acetone to minimize the drying time. The original and final masses of the ionic liquid and chromosorb mixture were in agreement to within 0.0005 g. Measurements were done at different temperatures between (303.15 and 373.15) K, i.e. (303, 313.15, 323.15, 333.15, 353.15, 363.15 and 373.15) K. Reproducibility was checked by undertaking three experimental runs. It was observed that retention times were reproducible within 0.05 minutes. The flow rate of dry helium, the carrier gas, was determined with the aid of a soap bubble flow meter placed at the outlet of the detector. The flow rates were corrected for water vapour pressure and varied from 0.3 to 0.7 um3.s-1 .The carrier gas flow rate was allowed to stabilize for at least 15 min prior to any series of runs. The pressure drop through the column varied from (25 to 50) kPa, providing conducive retention times and sharp peaks. The injected volumes ranged between 0.1 and 0.5 ul and were considered small enough to comply with the condition of infinite dilution of the solutes on the column. Both the injector and the detector were at T = 523.15 K. More details on the experimental parameters and variables needed to compute infinite dilution activity coefficients by means of equation (3-10), are discussed below.

4.1.3.1 Temperature control

The equipment, not only controls the temperature using a thermostat, but also displays its value

5

on the monitor. The column oven temperature was known with an accuracy of #177; 0.01 K and a stability of 0.05 K. The design of the GC was such that an electronically controlled thermostat was used to reliably control the column temperature.

4.1.3.2 Pressure measurement

The outlet pressure which is the same as the atmospheric pressure was measured with the aid of a digital barometer with an uncertainty of #177;0.30 %. The pressure drop through the packed column was fixed by the equipment, depending on the inert gas flow-rate set by the operator. The uncertainty in the inlet pressure measurement was #177;0.50 %.

4.1.3.3 Flow-rate measurement

A bubble soap flow-meter consisting of a 100 ml calibrated cylinder was used for the determination of helium flow rate. The uncertainty in the flow rate measurement was estimated as #177;0.20 %.

4.1.3.4 Infinite dilution range

Injections between 0.1 and 0.5ul were considered small enough to comply with the infinite dilution requirement. It is however advisable to check whether the retention time for a solute at the selected injection volume and flow rate remains the same as when it is part of a mixture.

4.1.3.5 Column packing

This is the most critical step when undertaking measurements via the Gas-Liquid Chromatography technique. Large error margins are caused by a careless column packing. The uncertainty in determining the mass of the solvent loaded into the column was #177;0.03 %.

Column cleaning

The column was washed with soapy water and flashed with acetone to facilitate drying with air.

Determination of solvent number of moles

The following procedure was employed:

1. The mass of an empty flask was measured using a digital balance;

2. Chromosorb, the solid support for the solvent was added to the flask and weighed;

3. The amount of solvent to be added was roughly calculated on the basis of the desired percent loading of the solvent;

4. The solvent was added and its weight found accurately.

Coating the support with the IL

Dichloromethane was added to the flask content to distribute the solvent evenly on chromosorb. Thereafter, a rotary evaporator was used to remove the Dichloromethane. This step ended when the mass of the mixture in the content was equal to the one measured before adding dichloromethane.

Filling the column

The support loaded with IL was filled in the column with the aid of a vacuum pump. One had to
make sure that the load was equally distributed inside the column. The mass of the packed

column had to be known before and after each series of runs to check for a probable elution of the solvent.

précédent sommaire suivant






Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy








"L'imagination est plus importante que le savoir"   Albert Einstein