WOW !! MUCH LOVE ! SO WORLD PEACE !
Fond bitcoin pour l'amélioration du site: 1memzGeKS7CB3ECNkzSn2qHwxU6NZoJ8o
  Dogecoin (tips/pourboires): DCLoo9Dd4qECqpMLurdgGnaoqbftj16Nvp


Home | Publier un mémoire | Une page au hasard

 > 

Contrainte Psycho-Physiques et Electrophysiologiques sur le codage de la stimulation électrique chez les sujets porteurs d'un implant cochléaire

( Télécharger le fichier original )
par Stéphane GALLEGO
Université Lyon I - Doctorat 1999
  

précédent sommaire suivant

Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy

- Article 18 :

LOUDNESS GROWTH FUNCTIONS AND EABR CHARACTERISTICS IN DIGISONIC COCHLEAR
IMPLANTEES

S. Gallégo, S. Garnier, C. Micheyl, E. Truy, A. Morgon, L. Collet
Acta Otolaryngol, 1999, 119, 234-238

L'objectif de cet article est de comparer la fonction de sonie aux caractéristiques des PEAEP.

Deux fréquences de stimulation sont analysées 75 et 300 Hz :

- Le 75 Hz correspond à une gamme compatible pour le recueil de PEAEP,

- Le 300 Hz correspond à la fréquence utilisée par le processeur pour coder le signal acoustique.

La mesure de la fonction de sonie est effectuée, par une technique catégorielle à 7 niveaux (très faible, faible, confortable mais faible, confortable, confortable mais fort, fort, très fort), sur une électrode apicale à deux fréquences de stimulation (75 et 300 Hz) chez 14 sujets implantés cochléaires Digisonic.

Le recueil des PEAEP s'effectue chez la même population, sur la même électrode à une fréquence de 75 Hz aux 7 niveaux de sonie calculés précédemment.

La comparaison entre les fonctions de sonie et la fonction entrée/sortie des PEAEP montre qu' :

Il est possible de déterminer très précisément le seuil de perception à 75 Hz et à 300 Hz à partir sur seuil d'apparition des PEAEP (respectivement, N=14, R2=0.9, p<0.001, N=14, R2=0.9, p<0.001),

Il est possible d'estimer le seuil d'inconfort à 300 Hz à partir de l'étude de la saturation de l'amplitude de l'onde V des PEAEP à 75 Hz,

L'amplitude de l'onde V évolue de manière linéaire du seuil de détection à 75 Hz jusqu'au seuil d'inconfort à 300 Hz.

La détermination de la fonction de sonie et plus particulièrement des seuils de détection et de confort à 300 Hz peuvent être estimer par l'analyse de l'onde V des recueils de PEAEP.

Acta Otolaryngol (Stockh) 1999; 119:

Loudness Growth Functions and EABR Characteristics in Digisonic Cochlear Implantees

S. GALLEGO
·, S. GARNIER', C. MICHEYL', E. TRUY
·, A. MORGON2 and L. COLLET"

From the 'CNRS UPRESA 5020 Laboratory and 2ENT Department, Par. U, Ed. Herriot Hospital. Lyon and the 311,IXM Laboratory, Vallauris, France

Gallego S, Garnier S, Micheyl C, Truy E, Morgon A, Collet L. Loudness growth functions and EABR characteristics in Digisonic cochlear implantées. Acta Otolaryngol (Stockh) 1999; 119: 000-000.

Electrically evoked auditory brainstem responses (EABRs) and loudness functions were measured in 14 subjects equipped with an MXM Digisonic cochlear implant. EABRs were evoked by 75-Hz pulse trains presented on the apical electrode. Loudness functions at the same rate and at a rate more conventional for psychoacoustic measurements (300 Hz) were measured using a categorical loudness-scaling procedure. The results revealed a significant difference in the loudness functions measured at 75 and 300 Hz, loudness increasing more steeply with stimulus intensity for the latter rate. Significant correlations between EABR wave V thresholds and perceptual thresholds measured at both 75 and 300 Hi were observed. Furthermore, in 8 out of the 14 patients, EABR wave V saturated at a stimulus level corresponding precisely to the loudest bearable, i.e. "Too loud" level for the 300-Hz stimulation rate: this same level corresponded to the "Comfortable" loudness level for the 75-Hz stimulation rate. On average, an almost linear relationship was observed over the first half of the loudness range between the stimulus intensity, expressed as a pulse duration in log units, and wave V amplitude in dB. Although further investigation is required before maximum comfort levels can be predicted reliably from EABR measures in individual subjects, these results indicate new directions regarding the estimation of perceptual dynamic range limits on the basis of EABR measures in cochlear implantees. Key words: Please supply Key Words.

pect EABR amplitude growth functions to be related to loudness growth functions. However, so far, no clear relationship has been established between EABR wave characteristics and UCLs or other aspects of the loudness functions in CI subjects.

In the present study, we measured the EABR amplitude growth functions and loudness growth functions in 14 subjects implanted with an MXM Digisonic DX10 cochlear implant. In order to facilitate comparisons between electrophysiological and psychoacoustical measurements, the same stimulus rate of 75 Hz was used in the two types of measures. However, because this 75-Hz rate is not generally used in normal conditions of implant functioning, we also measured loudness functions at a 300-Hz stimulation rate, which is more generally used in speech coding strategies with the Digisonic device.

MATERIAL AND METHODS Subjects

Fourteen subjects (6 female, 8 male) aged from 25 to 65 years, implanted with the Digisonic DX10, took part in the study. They were ail profoundly deaf before implantation.

The Digisonic DX10 cochlear implant

The Digisonic DX10 is a 15-electrode transcutaneous cochlear implant (4). The device comprises an external and an internai, i.e. implanted part (Fig. 1). The external device performs an analysis of the acoustic

INTRODUCTION

An important step in cochlear implant (CI) fitting consists in the measurement of the dynamic range of hearing in electric stimulation. This dynamic range is bounded on one side by the faintest stimulus intensity that the subject can perceive, or the threshold, and on the other side by the maximum intensity that the subject can bear, which is otherwise called the uncomfortable level (UCL). These lower and upper limits of the dynamic range depend on several factors and are highly variable across subjects. Thus, measuring these levels in the cochlear implant patient whose device is to be set often is a necessary preliminary step in the fitting process.

Thresholds and UCLs can be determined using psychophysical procedures in adults. However, these subjective procedures are generally time-consuming, especially when accurate measurements are needed. Furthermore, they cannot always be used in children. Therefore, they would advantageously be replaced by objective measures.

Data in the literature indicate the existence of relationships between electrophysiological variables and perceptual thresholds in CI subjects. In particular, absolute thresholds have recently been shown to be correlated to the thresholds of electrically evoked auditory brainstem response (EABR) waves (1-3). If EABR thresholds are correlated to perceptual thresholds, one might expect EABR saturation levels to be related to UCLs. Furthermore, one might ex-

Scandinavian University Press. ISSN 0001-6489

signal, which leads to the extraction of relevant speech information. This analysis mainly involves analog-to-digital conversion of the acoustic signal at a sampling rate of 15.6 kHz and real-time FFT computation over a 128-point analysis window. The resulting spectrum is composed of 64 frequency bands, 122 Hz wide each, within a 0-7.8 kHz range. The implanted part of the device performs the decoding of the information received by the internai antenna and stimulates the electrodes. With normal insertion of the electrode array into the cochlea, the 15 electrodes are typically located 5-20 mm from the base.

For EABR recordings and psychophysical measurements, the external part of the device was replaced by a dedicated stimulation system, MXM Digistim, which was piloted from a personal computer through an opto-isolated serial port. This system allowed the control of the stimulus parameters: stimulated electrode, pulse magnitude, pulse duration, and stimulation frequency. Synchronization between the stimulation and EABR-recording systems was âchieved using an external trigger pulse produced by the Digistim.

The stimulation mode used in this study was "Common Ground", wherein only the addressed electrodes are activated, ail the other electrodes being connected to ground (5). The electric stimuli were produced by a pulse generator, which was able to deliver 0.5-3 mA. The pulses were of constant amplitude, but variable duration. Pulse durations ranged between 5-500 sus. Pulses were delivered on the apical electrode. Two stimulation rates were set: 75 and 300 Hz. The duration of the train pulse was 400 ms.

Loudness growth function measurements

Loudness growth functions were measured using a loudness scaling procedure (6). Forty different stimulation levels, equally spaced within the dynamic range, i.e. between the threshold and discomfort levels, were presented to the subjects, three times, in random order. On each presentation, the subject's task was to rate the loudness using one of the following categories: (0) Inaudible, (1) Very soft, (2) Soft, (3) Comfortable but soft, (4) Comfortable, (5) Comfortable but loud, (6) Loud, (7) Very loud, (8) Too loud. Loudness functions were obtained by computing the average stimulus intensity corresponding to each loudness category.

EABR recordings

EABR recordings were performed according to a method described in previous publications (1, 3, 7). The electrodes were positioned according to an Fpz placement plan, the ground electrode being placed on the contralateral lobe. Full-scale sensitivity was set at 200 AT. Bandwidth was 10-3,000 Hz. The number of sweeps for trace averaging was set at 2,000. The analysis window duration was 10 ms, corresponding to 1,024 points. The stimulation rate was 75 Hz. Stimulus artefacts were cancelled out using adaptive digital filtering (8). Fig. 2 shows typical EABR recordings obtained using this method.

Microphone


·

Behind the ear device

Skin


·

Processing and
coding

MI

Implanted electrodes

CK:_){2{:}27.C
·DC}CliCaCD*1.--i

Internai
receptor

External
antenna

Fig. I. Block diagram of the DX10 Digisonic cochlear implant. Information is transmitted from the speech processor to the internai part on a carrier wave using amplitude modulation. The carrier frequency is 4 MHz.

PIII

PV

0.2 pV

1 ms

Pll

0 1

Fig. 2. Example of. EABR recording. Waves II, III and V are iden2 3 4 5 6 7 8 tified.

tween the perceptual thresholds at there two frequencies.

An important new result obtained in the present study is that in 8 out of 14 patients, EABR wave V amplitude saturated at a stimulus level corresponding precisely to the loudness level judged as "Too loud" by the subjects with the 300-Hz stimulation rate. This "Too loud" level at 300 Hz happened to correspond to the "Comfortable" level at 75 Hz. The relationships between stimulus intensity, loudness and wave V amplitude are represented in Fig. 6, with data averaged across the 14 subjects. It can be seen on this graph that over the first half of the loudness range, i.e. over the first four loudness levels at 75 Hz, there is an almost linear relationship between wave V

RESULTS AND DISCUSSION

Loudness growth functions

Fig. 3 shows functions relating pulse widths to loudness categories. The results of a two-way repeated measures ANOVA revealed a significant effect of stimulation rate (F 30.5, p < 0.001). Pulse widths corresponding to a given loudness level were generally larger at 75 Hz than at 300 Hz. An exception to this was that at the "Very soft" loudness level, which corresponded to threshold, post hoc tests revealed no significant effect of stimulation rate.

EABRs

--e40

025

20

2 3 4 5 6

Loudness scale

7

Fig. 4 shows EABR wave latencies as a function of stimulus loudness. Waves II, III and V average latencies are in agreement with data from the literature (9). As can be seen on the graph, EABR latencies were not very sensitive to the loudness. The results of two-way ANOVA indicated no significant effect of the loudness level on wave III and V latencies, but a significant effect on wave II latency (F = 4.651, p < 0.001). Fig. 5 shows EABR wave amplitudes as a function of stimulus loudness. The results of a twoway repeated measures ANOVA indicated a significant effect of loudness on the amplitude of waves II (F = 5.19, p < 0.001), III (F = 6.751 p < 0.001) and V (F = 4.049 p < 0.001).

Fig. 3. Functions relating stimulus intensity and loudness at two stimulation rates (75 and 300 Hz) averaged across the 14 cochlear implantees. The correspondence between numerical loudness scores on the abscissa and verbal loudness categories is as follows: 1 Very soft, 2 = Soft, 3 = Comfortable but soft, 4 = Comfortable, 5 = Comfortable but loud, 6 = Loud, 7 = Very loud. Stimulus intensities are expressed as pulse widths in dB units, i.e. 201og10 (pulse width in ps) on the ordinate. The error bars represent the standard error of the mean.

In agreement with previous results (1-3), significant correlations were observed between EABR wave V thresholds and perceptual thresholds measured at 75 and 300 Hz (Pearson's r 0.9 p < 0.001, n= 14). This finding, that EABR thresholds correlated with perceptual thresholds measured using the same stimulus rate (75 Hz), but also a different rate (300 Hz) can be explained by the above-mentioned similarity be-

4.5
4.0
g 3.5

c; 3.0 a.)

-5 2.5

o 2.0

1.5

 

2 3 4 5 6 7

Loudness scale

Fig. 4. Functions relating wave II, HI and V latencies to loudness. The scale on the abscissa is the same as in Fig. 3. Data are averaged across the 14 subjects; the error bars represent the standard error of the mean.

amplitude in dB and stimulus intensity, i.e. pulse width in log units. As can be seen on the graph, on average, the stimulus level at which wave V amplitude saturated corresponded to the "Comfortable" loudness level at 75 Hz and to the "Too loud" or uncomfortable loudness level (UCL) at 300 Hz. This relationship did not hold over the upper half of the loudness rate (beyond the "Comfortable" loudness level) due to saturation of wave V amplitude.

CONCLUSION

The present results confirm and complement previous reports indicating the existence of statistical relationships between EABR and perceptual thresholds in subjects implanted with the MXM Digisonic cochlear implant. A new element brought forward by the present study is that EABR wave V amplitude obtained at a 75-Hz stimulation rate allows the prediction of not only the absolute thresholds at this and

,-..., 60

cip

-a 50

a.) -a

40

E

° 30

o

e 20

2 3 4 5 6 7

Loudness scale

Fig. 5. Functions relating wave II, III and V amplitudes to loudness. The scale on the abscissa is the same as in Fig. 4. Data are averaged across the 14 subjects; the error bars represent the standard error of the mean.

Fig. 6. Relationship between stimulus intensity, loudness, and EABR wave V amplitude. The different loudness categories are represented by different symbols. The graph is based on data obtained using a 75-Hz stimulation rate, averaged across the 14 subjects; the error bars represent the standard error of the mean. The dotted lines correspond to the threshold (left) and discomfort level (right) for the 300-Hz stimulation rate.

TL (300 HZ)

UCL (300 HZ)

25 30 35

Stimulus intensityy (dB 1.1.$)

40

other frequencies, but also other components of the loudness range. These include the following: the stimulus level deemed "Comfortable" by the subjects at the same low rate (75 Hz) or the "Uncomfortable" level measured with a higher rate (300 Hz), which is more conventionally used to convey information to the auditory system of cochlear implantees. Further work is certainly required before the statistical relationships between the electrophysiological and perceptual variables evidenced here can be used at the individual level for the fitting of the device.

ACKNOWLEDGMENTS

The authors acknowledge the support of the MXM Company, the ANRT, the Hospices Civils of Lyon, the CNRS, the University Lyon I, and the cochlear implantees who took part in the study.

REFERENCES

I. Gallégo S, Truy E, Morgon A, Collet L. EABRs and

surface potentials with a transcutaneous multielectrode

cochlear implant. Acta Otolaryngol (Stockh) 1997; 117:

164-8.

2. Truy E, Gallégo S, Chanal JM, Collet L, Morgon A. Correlation between electrical ABR and perceptual thresholds in Digisonic cochlear implants users. Laryngoscope 1997; 118: 554-9.

3. Gallégo S, Frachet B, Micheyl C, Truy E, Collet L. Cochlear implant performance and electrically auditory brainstem response characteristics. Electroencephalogr Clin Neurophysiol. In press.

4. Beliaeff M, Dubus P, Leveau JM, Repetto JC, Vincent P. Sound processing and stimulation coding of Digisonic DX10 15-channel cochlear implant. In: Hochmair IN, ed. Advances in cochlear implant. Innsbruck:

Verlag, 1994: 198-203.

5. Gallégo S, Ba Lê L, Berger-Vachon C. Modelling of the electrical stimulation delivered by the Digisonic Multichannel cochlear implant. Adv Model Anal 1998; 53: 1-15.

6. Jerger H, Olivier TA, Chmiel RA. Prediction of dynamic range from stapedius reflex in cochlear implanted patients. Ear Hear 1988; 9: 4-8.

7. Gallégo S, Micheyl C, Berger-Vachon C, Truy E, Morgon A, Collet L. Ipsilateral ABR with cochlear implant. Acta Otolaryngol (Stockh) 1996; 116: 228-33.

8. Gallégo S, Collet L, Berger-Vachon C. Electrically auditory brainstem responses (EABR): contribution of a

filter adapted to the auditory system. World Congress on Medical Physics and Biomedical Engineering. Nice, 14-19 September 1997. Med Biol Eng Comput 1997; 35 (Suppl 1): 304.

9. Abbas PJ, Brown CJ. Electrically evoked brainstem potentials in cochlear implant patients with multi-electrode stimulation. Hear Res 1988; 36: 53-162.

Address for correspondence:

Stéphane Gallégo

Sté MXM, 2720 chemin Saint Bernard 06227 Vallauris Cedex

France

Tel: + 33 4 93 95 18 18

Fax: + 33 4 93 95 38 01

E-mail: s2allego@mxmlab.com

cl PEAEP et discrimination d'intensité

La mesure du seuil différentiel en intensité est autant voir même plus important que la mesure de sonie, car elle va déterminer la résolution du codage aux différents niveaux en intensité (cf chapitre dl, enveloppe).

Des travaux chez le normo-entendant, n'ont pas pu reconstituer la fonction de sonie à partir de la mesure de discrimination en intensité car les mécanismes qui codent l'intensité de stimulation sont multiples et non linéaires.

Des mesures de la sonie et du seuil différentiel en intensité en fonction de la durée du pulse effectuées sur le Digisonic n'ont pas montré d'irrégularités, ce qui n'est pas le cas chez le normo-entendant (Raab et Taub, 1969). La détermination des IDL a partir des PEAEP chez les sujets implantés cochléaires semble donc possible.

précédent sommaire suivant






Bitcoin is a swarm of cyber hornets serving the goddess of wisdom, feeding on the fire of truth, exponentially growing ever smarter, faster, and stronger behind a wall of encrypted energy








"I don't believe we shall ever have a good money again before we take the thing out of the hand of governments. We can't take it violently, out of the hands of governments, all we can do is by some sly roundabout way introduce something that they can't stop ..."   Friedrich Hayek (1899-1992) en 1984